Current issue: 58(5)

Scopus CiteScore 2023: 3.5
Scopus ranking of open access forestry journals: 17th
PlanS compliant
Select issue
Silva Fennica 1926-1997
1990-1997
1980-1989
1970-1979
1960-1969
Acta Forestalia Fennica
1953-1968
1933-1952
1913-1932

Articles by Jyrki Hari

Category : Article

article id 5012, category Article
Jyrki Hari, Pertti Hari, Markku Kanninen. (1978). An electronic auxanometer for field use. Silva Fennica vol. 12 no. 4 article id 5012. https://doi.org/10.14214/sf.a14866
Keywords: diameter; diameter growth; measuring instruments
Abstract | View details | Full text in PDF | Author Info

This paper describes the design and functioning of an electronic auxanometer designed for field use, and some results obtained with it. The changes in stem radius are monitored with a micrometer screw turned by an electirc motor. A potentiometer transforms the position of the screw into an electronic signal, which is recorded. The accuracy of the device is approximately 1–2 μm. Field experiments with the auxanometer are described and discussed. The accuracy of the device was found to be sufficient for measuring hourly changes in stem radius.

The PDF includes a summary in Finnish.

  • Hari, E-mail: jh@mm.unknown (email)
  • Hari, E-mail: ph@mm.unknown
  • Kanninen, E-mail: mk@mm.unknown

Register
Click this link to register to Silva Fennica.
Log in
If you are a registered user, log in to save your selected articles for later access.
Contents alert
Sign up to receive alerts of new content
Your selected articles